Philips-Respironics error codes
This is a listing of CPAP Error Codes and descriptions for Philips-Respironics 50-series and 60-series System One REMstar and BiPAP machines.
We do have the Service & Repair manuals for these and other CPAP machines available. To access them, you will need to join the Apnea Board Forum, log in, and then go to our Private Files and Links area, below the forums. In that forum, there is a post that says "Links to Members-Only Files". If you open that post, you can click on our current link to the Members Only Files. Please keep in mind that we do not have additional Service/Repair Manuals other than those listed on our Members Only Files page.
Contents
- 1 Types of Error Codes
- 2 General Errors
- 3 Watchdog and Timer Errors
- 4 Motor / Blower Errors
- 5 NVRAM Low Level Errors
- 6 NVRAM Unit Related Errors
- 7 General Hardware Errors
- 8 Pressure Sensor Errors
- 9 Flow Sensor Errors
- 10 OTP and RTC Errors
- 11 HUMIDIFIER Errors
- 12 Stack and Exception Handler Errors
- 13 See Also
Types of Error Codes
The four Error Types are described as follows:
Error Type | Description |
---|---|
STOP | The error information is recorded in NVRAM and the unit is placed into Safe State. The only functionality available to the user is serial communication, turning off the audible alarm via a key press and removing power. |
REBOOT | The error information is recorded in NVRAM and the unit is rebooted. This error is promoted to a STOP level error by use of the Verbose Mode configuration flag. |
CONTINUE | The error information is recorded in NVRAM and the unit continues to operate without noticeable alteration. This error is promoted to a STOP level error by use of the Verbose Mode configuration flag. |
LOG_ONLY | The error information is recorded in NVRAM and the unit continues to operate without noticeable alteration. This error is not promoted to a STOP level error by use of the Verbose Mode configuration flag. |
General Errors
Error Code | Description | Type |
---|---|---|
E-1 | ERR_SOFTWARE_STOP | STOP |
E-2 | Not Used | N/A |
E-3 | ERR_INT_RAM | REBOOT |
E-4 | ERR_NULL_PTR | REBOOT |
E-5 | ERR_DATA | REBOOT |
E-6 | ERR_STATE_MACHINE | REBOOT |
E-7 | ERR_SOFTWARE | REBOOT |
E-8 | Not Used | N/A |
E-9 | Not Used | N/A |
Watchdog and Timer Errors
Error Code | Description | Type |
---|---|---|
E-10 | ERR_WDOG_TEST_RAM | REBOOT |
E-11 | ERR_WDOG_TEST | REBOOT |
E-12 | ERR_BACKGROUND_WDOG_NO_CARD | REBOOT |
E-13 | ERR_BACKGROUND_WDOG_SD_CARD | REBOOT |
E-14 | ERR_WDOG_LOWRES_TIMER | REBOOT |
E-15 | ERR_CYCLE_HANDLER_OVERRUN | REBOOT |
E-16 | ERR_RASP_RESTORE_TIMEOUT | CONTINUE |
E-17 | ERR_ONEMS_HANDLER_OVERRUN | REBOOT |
E-18 | Not Used | N/A |
E-19 | ERR_WDOG_TIMEOUT | REBOOT |
Motor / Blower Errors
Error Code | Description | Type |
---|---|---|
E-20 | ERR_MOTOR_SPINUP_FLUX_LOW | REBOOT |
E-21 | ERR_MOTOR_VBUS_HIGH | STOP |
E-22 | ERR_MOTOR_FLUX_MAGNITUDE | REBOOT |
E-23 | ERR_MOTOR_OVERSPEED | REBOOT |
E-24 | ERR_MOTOR_SPEED_REVERSE | REBOOT |
E-25 | ERR_MOTOR_THERMISTOR_OPEN | CONTINUE |
E-26 | ERR_MOTOR_THERMISTOR_SHORTED | CONTINUE |
E-27 | ERR_MOTOR_RL_NOCONVERGE | STOP |
E-28 | ERR_NEGATIVE_QUADRATURE_VOLTAGE | REBOOT |
E-29 | ERR_VBUS_GAIN_ZERO: | REBOOT |
E-30 | ERR_MOTOR_SPINUP_FLUX_HIGH | REBOOT |
E-31 thru E-39 | Not Used | N/A |
NVRAM Low Level Errors
Error Code | Description | Type |
---|---|---|
E-40 | ERR_NVRAM | REBOOT |
E-41 | ERR_STORAGE_UNIT_RAM | REBOOT |
E-42 | ERR_UNABLE_TO_OBTAIN_BUS | REBOOT |
E-43 | ERR_NVRAM_NO_CALLBACK_OCCURRED | REBOOT |
E-44 | ERR_NV_BUFFER_NULL | REBOOT |
E-45 | ERR_NV_CALLBACK_NULL | REBOOT |
E-46 | ERR_NV_ZERO_LENGTH | REBOOT |
E-47 | ERR_NVRAM_INVALID_BYTES_XFRRED | REBOOT |
E-48 | Not Used | N/A |
E-49 | Not Used | N/A |
NVRAM Unit Related Errors
Error Code | Description | Type |
---|---|---|
E-50 | ERR_DAILY_VALUES_CORRUPT | LOG_ONLY |
E-51 | ERR_CORRUPT_COMPLIANCE_LOG | CONTINUE |
E-52 | ERR_CORRUPT_COMPLIANCE_CB | CONTINUE |
E-53 | ERR_COMP_LOG_SEM_TIMEOUT | CONTINUE |
E-54 | ERR_COMPLOG_REQS_OVERFLOW | REBOOT |
E-55 | ERR_THERAPY_QUEUE_FULL | CONTINUE |
E-56 | ERR_COMPLOG_PACKET_STATUS | REBOOT |
E-57 | ERR_SESS_OBS_QUEUE_OVF | REBOOT |
E-58 | ERR_SESS_OBS_NO_CALLBACK | REBOOT |
E-59 | Not Used | N/A |
General Hardware Errors
Error Code | Description | Type |
---|---|---|
E-60 | ERR_UNSUPPORTED_HARDWARE | REBOOT |
E-61 | ERR_PLL_UNLOCKED | REBOOT |
E-62 | ERR_STUCK_RAMP_KEY | CONTINUE |
E-63 | ERR_STUCK_KNOB_KEY | CONTINUE |
E-64 | ERR_DSP_OVERTIME_PWM | REBOOT |
E-65 | ERR_STUCK_ENCODER_A | CONTINUE |
E-66 | ERR_STUCK_ENCODER_B | CONTINUE |
E-67 | Not Used | N/A |
E-68 | Not Used | N/A |
E-69 | Not Used | N/A |
Pressure Sensor Errors
Error Code | Description | Type |
---|---|---|
E-70 | ERR_PRESSURE_SENSOR_ABSENT | STOP |
E-71 | Not Used | N/A |
E-72 | ERR_PSENS_UNABLE_TO_OBTAIN_BUS | REBOOT |
E-73 | ERR_SENSOR_PRESS_OFFSET_STOP | STOP |
E-74 | Not Used | N/A |
E-75 | Not Used | N/A |
E-76 | Not Used | N/A |
E-77 | Not Used | N/A |
E-78 | Not Used | N/A |
E-79 | Not Used | N/A |
Flow Sensor Errors
Error Code | Description | Type |
---|---|---|
E-80 | ERR_UNABLE_TO_INIT_FLOW_SENSOR | REBOOT |
E-81 | ERR__FLOW_SENSOR_TABLE | CONTINUE |
E-82 | ERR_FLOW_SENSOR_OFFSET | CONTINUE |
E-83 | ERR_FSENS_UNABLE_TO_OBTAIN_BUS | REBOOT * |
E-84 | ERR_FLOW_SENSOR_ STOP | STOP |
E-85 | ERR_FLOW_SENSOR_OCCLUDED | CONTINUE |
E-86 | ERR_FLOW_SENSOR_ABSENT | CONTINUE |
E-87 | ERR_FLOW_SENSOR_BUS | CONTINUE |
E-88 | Not Used | N/A |
E-89 | Not Used | N/A |
*On E-83, upon reboot, a second failure results in a STOP level error.
OTP and RTC Errors
Error Code | Description | Type |
---|---|---|
E-90 | ERR_OTP_NOT_CONFIGURED | STOP |
E-91 | ERR_OTP_INCORRECTLY_CONFIGURED | STOP |
E-92 | Not Used | N/A |
E-93 | ERR_RTC_VALUE | CONTINUE |
E-94 | ERR_RTC_STOPPED | CONTINUE |
E-95 thru E-99 | Not Used | N/A |
HUMIDIFIER Errors
Error Code | Description | Type |
---|---|---|
E-100 | ERR_HUMID_NO_HEAT | CONTINUE |
E-101 | ERR_HUMID_TEMP_MAX | STOP |
E-102 | ERR_THERMISTOR_HIGH | CONTINUE |
E-103 | ERR_THERMISTOR_LOW | CONTINUE |
E-104 | ERR_HUMID_AMBIENT_OFF | CONTINUE |
E-105 | ERR_HUMID_AMBIENT_COMM | CONTINUE |
E-106 thru E-109 | Not Used | N/A |
Stack and Exception Handler Errors
Error Code | Description | Type |
---|---|---|
E-110 | ERR_STACK | REBOOT |
E-111 | ERR_EXCEPTION_STACK_OVERFLOW | REBOOT |
E-112 | ERR_EXCEPTION_STACK_RESERVE | LOG_ONLY |
E-113 | ERR_EXCEPTION_STACK_UNDERFLOW | REBOOT |
E-114 | ERR_FIQ_STACK_OVERFLOW | REBOOT |
E-115 | ERR_FIQ_STACK_RESERVE | LOG_ONLY |
E-116 | ERR_FIQ_STACK_UNDERFLOW | REBOOT |
E-117 | ERR_IRQ_STACK_OVERFLOW | REBOOT |
E-118 | ERR_IRQ_STACK_RESERVE | LOG_ONLY |
E-119 | ERR_IRQ_STACK_UNDERFLOW | REBOOT |
E-120 | ERR_SVC_STACK_OVERFLOW | REBOOT |
E-121 | ERR_SVC_STACK_RESERVE | LOG_ONLY |
E-122 | ERR_SVC_STACK_UNDERFLOW | REBOOT |
E-123 | ERR_DATA_ABORT_EXCEPTION | REBOOT |
E-124 | ERR_PREFETCH_EXCEPTION | REBOOT |
E-125 | ERR_SWI_ABORT_EXCEPTION | REBOOT |
E-126 | Not Used | REBOOT |
See Also
Philips-Respironics DreamStation error codes
REMstar error codes (for older, pre-2007 REMstars)
REMstar smart card error codes
Donate to Apnea Board